Power Range | Power Range +10.0 to + 30.0 D in 0.5 D (+5.0 to +9.0 and +31.0 to +40.0 in 1.0 D) increments |
Optic Body Diameter | 6.0 mm |
Overall length with haptics | 13.0 mm |
Design | Single piece refractive optic |
Material | UV Absorbing Hydrophilic Acrylic Co-polymer (From : Contamac Ltd., U.K.) |
Edge design | 360º posterior square edge |
Suggested A-Constant* | Optical : 117.7, Ultrasonic 117.5 (Select first minus) |
Theoretical AC- depth* | 4.96 mm |
Surgeon / Lens Factor* | 1.73 |
Design Factor | +6.0 D |
Incision Size | 2.4 mm or more recommended, please never attempt wound assisted delivery |
A-constant, ACD, and surgeon / lens factor are theoretical estimates only. It is recommended that each surgeon develop his / her own values.
INDICATIONS: Indicated for primary implantation for the visual correction of aphakia in patients in whom the cataractous lens has been removed by an extra-capsular cataract
extraction method. The lens is intended for placement in the capsular bag. Can also be implanted to correct presbyopia in appropriate spaces.
PRECAUTIONS: Physicians considering lens implantation under any of the following circumstances should weigh the potential risk/benefit ratio:
1. Recurrent severe anterior or
posterior segment inflammation or uveitis.
2. Patients in whom the intra ocular lens may affect the ability to observe, diagnose, or treat posterior segment diseases.
3. Surgical
difficulties at the time of cataract extraction, which might increase the potential for complications (eg, persistent bleeding, significant iris damage, uncontrolled positive pressure, or
significant vitreous prolapse or loss).
4. A distorted eye due to previous trauma or developmental defect in which appropriate support of the IOL is not possible.
5. Circumstances that
would result in damage to the endothelium during implantation.
6. Suspected microbial infection.
7. Children under the age of 2 years are not suitable candidates for intra ocular
lenses.
8. Patients in whom either the posterior capsule or zonules are not intact enough to provide support.
WARNINGS: Do not attempt to re-sterilise the lens. Do not soak or rinse the intra ocular lens with any solution other than sterile balanced salt solution, sterile Ringer’s lactate or sterile
normal saline. Do not store the lens at a temperature greater than 43°C (110°F). DO NOT FREEZE. Do not autoclave the intra ocular lens. Do not reuse the lens. It is intended for
permanent implantation.